Eagle Home Products Eagle Series instruction manual Adjusting the sensor controls

Page 62

Adjusting the sensor controls

CAUTION:Only perform the following steps if you are unable to set the correct calibration reading with the AIR/and SHIFT/buttons.

1.Use the AIR/and SHIFT/buttons to set the reading to the middle of the range in which you can currently adjust the reading. For example, if you can set the reading from a minimum of 1.0 ppm to a maximum of 3.0 ppm, set the display to 2.0 ppm.

2.Unscrew the two large screws on the top of the case, then carefully lift the top of the case and lay it aside.

3.Locate the sensor in the front half of the bottom case. The sensor with wires connected to EC3 on the analog PCB displays its reading on the third line of the screen. The sensor connected to EC4 on the analog PCB displays its reading on the fourth line of the screen.

4.Adjust the sensor’s SPAN control one turn and observe the display reading. Continue to adjust the SPAN control until the display reading matches the concentration of the calibration cylinder.

The SPAN control is located next to the sensor on the side closest to the front of the instrument.

5.Press the POWER/ENTER button to set the span value. SINGLE CALIBRATION END displays, then the Single Calibration menu displays.

6.Place the top of the case in its original position, then secure it with the large screws you loosened in step 2.

7.Disconnect the tubing from the probe, then unscrew the regulator from the calibration cylinder.

8.With the Single Calibration menu displayed, press the SHIFT/button until the prompt is next to the last channel, then press the SHIFT/button again. The ESCAPE message displays.

9.Press the POWER/ENTER button to return to the Calibration menu.

10.Press the SHIFT/button to place the prompt next to the NORMAL OPERATION menu option, then press the POWER/ ENTER button to return to the normal screen.

56 • Appendix C: Non-Standard Toxic Gas Sensors

Eagle Instruction Manual

Image 62 Contents
Portable Multi-Gas Detector RKI Instruments, Inc Whipple Rd. Hayward, CA 94544 510Warranty WarrantyTable of Contents Display Mode Table of ContentsCalibration Appendix F Eagle Tank Tester Model Introduction IntroductionOverview About this Manual Eagle Specifications SpecificationsSpecifications Carbon Standard Sensor SpecificationsCombustible Oxygen Control Panel DescriptionCase ButtonsInterface Port Alarm LightsBattery Charger Connector BuzzerSensors Circuit Boards Methane Elimination SwitchCAL/SETUP Switch Operation OperationStarting Up the Eagle Operation Normal Operation Monitoring Combustible Gas in the PPM RangeTarget Gas LEL Factor Full Response Mode Conversion Factors Methane CalibrationMonitoring Combustible Gases Other than Methane Setting User Access Turning Off the EagleAlarms AlarmsAlarm Indications Stel alarm toxics only Low flow alarm Sensor failure alarm and emergency operation Resetting Gas Alarms User and Station ID Screen Display ModeDisplay Mode To enter a user and station ID Peak ScreenElapsed Time Screen Date/Time Screen Battery Voltage ScreenTWA/STEL Screen To clear the data log Clear Data Logger ScreensRemaining Log Time Screen Setup Mode Setup ModeTips for Using Setup Mode Entering Setup Mode Updating the Battery Type SettingUpdating Channel Settings Setting a custom target gas label Fullscale PPM Readings Equivalent to 100% LEL Target GasUpdating oxygen channel settings Updating toxics channel settings Setting a custom target gas label Updating Combustible Gas Channel Units of Measure Updating the Alarm Point Settings Updating the Lunch Break Setting Updating the Eagle’s Serial NumberUpdating the Alarm Latching Setting Updating the Alarm Silence SettingUpdating the Auto Calibration Settings Turning the User ID Function On or OffUpdating the Back Light Setting Turning the Auto Fresh Air Function On or OffUpdating the Interval Time Setting data log option Updating Log Data Over Write Setting data log optionUpdating the Time Calibration Setting data log option Updating the Date and Time Settings data log optionReturning to Default Settings Updating the Zero Follow SettingsUpdating the Confirmation Beep Setting To reset the oxygen zero setting To reset all default settingsTo reset all default alarm point settings Preparing for Calibration CalibrationCalibration Supplies and Equipment CalibrationCalibrating the Eagle Calibration Calibrating with the Single Calibration method Calibration To display the battery voltage MaintenanceDisplaying the Battery Voltage Replacing Alkaline BatteriesReplacing Sensors To replace the combustibles sensorTo replace the oxygen sensor To replace the H2S or CO sensor Appendix a Parts List Appendix a Parts ListParts List Part Number DescriptionPart Number Appendix B Methane Elimination Appendix B Methane EliminationMonitoring combustible gases other than hexane Description Appendix C Non-Standard Toxic Gas SensorsSpecifications Non-Standard Toxic Gas Sensors SpecificationsCalibrating Non-Standard Toxic Gas Sensors Adjusting the sensor controls Replacing Non-Standard Toxic Gas Sensors Parts List Parts List Non-Standard Toxic Gas SensorsCarbon Dioxide Sensor Specifications Alarm TWA Alarm Stel AlarmAppendix D Carbon Dioxide Sensors Appendix D Carbon Dioxide SensorsCalibrating Carbon Dioxide Sensors Normal Operation of Carbon Dioxide SensorsDemand Zero for Carbon Dioxide Sensors Carbon Dioxide Fresh Air ReadingsAdjusting the sensor controls Replacing Carbon Dioxide Sensors Parts List Carbon Dioxide Sensors Target Gases Range Alarm TWA Alarm Stel AlarmAppendix E Infrared Methane Sensors Infrared Methane Sensor SpecificationsCalibrating Infrared Methane Sensors Sensor Analog PCB SocketParts List Infrared Methane Sensors Replacing Infrared Methane SensorsTo use the float probe assembly Appendix F Eagle Tank Tester ModelTo attach the dilution fitting AlarmsTo eliminate an oxygen alarm that is not being used Appendix F Eagle Tank Tester ModelParts List Eagle Tank Tester Model CalibrationAppendix G Five-Gas and Six-Gas Models Appendix G Five-Gas and Six-Gas ModelsDisplaying Additional Channels Calibration, Display, and Setup Modes Appendix H Eagle Transformer Gas Tester Model Appendix H Eagle Transformer Gas Tester ModelAlarms Appendix I Installing the Data Logger Board Appendix I Installing the Data Logger Board