K-Tec MT2000 manual Detailed Configuration Parameters Direct Mode Blanking Menu

Page 12

4.4Detailed Configuration Parameters

4.4.1Direct Mode Blanking Menu

Note: This section only applies to direct mode measurement.

The blanking parameter (BLK) is used to ignore an extended nozzle that would otherwise cause a reflected signal at the top of the probe and result in a high level reading even when no product is in the vessel.

For probe configuration 1,2,5,6 or 7 (Appendix G) use the factory default BLK.

BLK can be changed by accessing SET followed by the CFG menu (refer to menu chart). If the MT2000 is installed in an extended nozzle or other configuration where the top of the tank is further than 1 inch from the probe coupler set BLK as follows:

For probe length < 100 ft. (30.5 m) add 4 to the BLK value for every 1 in (2.5 cm) of tank nozzle For probe length ≥ 100 ft. (30.5 m) add 2 to the BLK value for every 1 in (2.5 cm) of tank nozzle.

Example: RS=1: Nozzle length = 10” BLK=200 + 4(10) = 240

To verify settings are correct, proceed to CFG menu. Scroll down to RC (Raw Count) value display. RC (Raw Count) will decrease as levels rises and increase as level lowers. Caution: if RC value exceeds 4 digits, then the first digit is shown on the top line of the display.

If RC stays at a fixed low value around 2000 while the level changes the BLK is set too low and needs to be ad- justed. Raise the level and stop when RC stops decreasing or top of the vessel is reached, whichever comes first. If RC stops decreasing before the top level is reached reduce the BLK.

To determine the actual level using Raw Counts use the following formula: (RC – 2100) / 38 = Distance in inches

(Distance x 38) + 2100 = Expected Raw Counts (RC) Typical BLK Factory settings:

Transmitter Configuration from Model Number

Range

/L Local Option

 

 

Probe length < 100 ft. / 30.5 m

BLK = 210

 

 

Probe length ≥ 100 ft. / 30.5 m

BLK = 110

Figure 4.3

4.4.2Threshold Level Parameter (THV) and Gain Setting (GS)

The threshold level and gain settings are used to fine tune the detection of products with different dielectric con- stants, and to adapt to different probe configurations. THV is typically set between 1.0 to 1.8.

Note: Lower THV values make the transmitter less sensitive and higher THV values make the unit more sensitive to lower dielectrics. GS is typically set to 2 or 4. GS = 1 is lower gain than GS = 2, and GS = 3 is lower Gain than GS = 4.

For probe configurations 1,2,3a, and 3b (Appendix G), with a dielectric constant above 10 use the factory default settings THV=1.5, GS=2.

For probe configurations 4,5,6,7,8,9,10a, 10b, THV settings of 1.65 ± 0.15 are typical, depending on fluid dielectric constant using GS = 4.

12

MT2000-0200-01 Rev f (10-2007) DCN0160

Image 12
Contents Guided Wave Radar Level Transmitter Table of Contents List of Figures AppendicesIntroduction Ambient Temperature OverviewStorage Information Measurement cycle consists of the followingDirect Reflect Mode standard cont’d Ultra Low Dielectric ULD Measure MethodInstallation Mounting RequirementsFactory Default Configuration Shortening of ProbeMT2000 Guided Wave Radar Guidelines URV = Probe Length LRV = 0 in. / 0 mmWiring Quickstart ProcedureCommissioning MT2000 Guided Wave Radar Guidelines cont’dQuickstart Procedure cont’d Verify Proper Power-UpSetting the 4mA and 20 mA Points Using the Pushbutton Menu LCD Menu Operation Calibration MenuDetailed Configuration Parameters Direct Mode Blanking Menu Threshold Level Parameter THV and Gain Setting GSThreshold Level Parameter THV and Gain Setting GS cont’d Advanced Parameter SettingsL1O Advanced Parameter Settings cont’d KO & KGUltra Low Dielectric ULD Mode Configuration = H * Sqrt DC H + LDC = I + H L / H2 Ultra Low Dielectric ULD Mode Configuration cont’d Advanced Output ConfigurationSetting the DAC trim DAC TrimFigure6 4mA & 20mA Calibration Using Actual Level Input Bench Calibration of the 4mA & 20mA PointsReversing The Output Action Using Actual Level Input Jumper Switch SettingsTroubleshooting Information Possible SymptomsValid Current Loop Outputs Possible Symptoms cont’d Electronics Module ReplacementHoneywell DE Output Option Interoperability and Conformance ClassHart Protocol Interface Option Operating ModesHart Communicator CompatibilityGlossary of Terms LRVAppendices Appendix a Sensor TrimConfiguration 2 Hidden Menu END CFG MenuTMP To access the linearization table Setting up Linearization Table PointsAppendix C Linearization To clear the linearization tableAppendix D Interface Applications MT2000 InterfaceUse and setup of the oscilloscope for the MT2000 Scope Settings Channel aPrerequisites Channel BAppendix F CE Certificate of Conformity Appendix G Exida Certificate MT2000-0200-1 Rev f 10-2007 DCN0160 MT2000-0200-01 Rev f 10-2007 DCN0160 MT2000-0200-1 Rev f 10-2007 DCN0160 Appendix I Mounting Configurations 3a, 3b10a 10bAppendix J Warranty Statement Year Warranty for