4.4.2Threshold Level Parameter (THV) and Gain Setting (GS) (cont’d)
The following values should be used as starting points:
For Mounting Configuration 1,2,3a and 3b, (direct insertion in tank - no stilling well | GS | THV | |
applications, no coaxial probes) with the following dielectric constant: | |||
|
| ||
|
|
| |
≥ 10 | 2* | 1.5* | |
|
|
| |
≥ 3 and < 10 | 2 | 1.7 | |
|
|
| |
|
|
| |
For Mounting Configuration 4,5,6,7,8,9,10a & 10b with the following dielectric |
|
| |
constant: (ground plane probe / installation applications - coaxial probes stilling | GS | THV | |
well, EC chambers) |
|
| |
≥ 10 | 4 | 1.0 | |
|
|
| |
< 10 | 4* | 1.7* | |
|
|
|
*Factory Default
Assuming the 4 mA point is set toward the bottom end of the probe and that the 20 mA point is set toward the top, for a given GS setting:
•If THV is set too high the output will either read high or spike high.
•If THV is set too low the output will either read low down to 3.8 mA, spike low, or be in fault high (21mA) or fault low (3.6mA).
•Adjust THV as needed to obtain a steady output reflecting the actual level over the usable measuring range.
NOTE: If a stable output cannot be obtained, refer to
4.4.3Advanced Parameter Settings
•L1O
Level One Offset is used to offset the level output of the transmitter to match the actual level in your tank or vessel. This offset allows the transmitter to be calibrated so that the 4mA is equal to a value higher than zero. There are several cases when Level Offset can be used:
1.To accommodate for a
2.To accommodate for the length of the probe being shorter than the ac- tual length of the tank or vessel due to agitator or other devices near the bottom.
3.To synchronize two redundant transmitters
4.To allow a
L1O can be positive or negative and as much as 50% of the sensor trim span (HTP- LTP). Figure 4.4 shows a typical L1O offset.
Example: If the end of the probe, in Figure
Measured
L1O
Figure 4.4
13 |