K-Tec MT2000 manual Troubleshooting Information, Valid Current Loop Outputs, Possible Symptoms

Page 19

5. TROUBLESHOOTING INFORMATION

Use a milliamp (mA) meter to measure the output current. When power is applied the output will go to 4.00 mA for at least one (1) second and then to either the measured level or an alarm condition output. If this does not happen the transmitter may not be receiving enough power or the main electronic is defective. Excessive current above 21.00 mA is also an indication of improper power up or defective electronics.

5.1Valid Current Loop Outputs

21.0 mA

If the top board jumper is set to HI ALARM a loss of signal, a

 

problem with the configuration or a malfunction will cause the

 

output to be set to the alarm condition of 21.0 mA

20.6 mA

When the level increases above the 20.0 mA point the output

 

will continue up to 20.6 mA and then latch until the level returns

 

below the 20.6 mA level.

4.0 mA to 20.0 mA

Normal output range between 3.8 and 20.6 mA

3.8 mA / (4.0 mA)

When the level decreases below the 4.0 mA point the output will

 

continue down to 3.8 mA and then latch until the level returns

 

above the 3.8 mA level. [If LRS (Low Range Saturation) is set

 

for 4.0 mA, that will be the lowest output, regardless of actual

 

level.]

3.6 mA

If the top board jumper is set to LO ALARM a loss of signal,

 

problem with the configuration, or a malfunction will cause the

 

output to be set to the alarm condition of 3.6 mA

Output spiking up or down will usually result from improper THV or BLK settings. Refer to the initial setup section for details. If a module fails to communicate, verify from the model number that the particular communication option is included.

Note: Signal can be observed using an oscilloscope. Refer to Appendix F for setup procedure.

5.2

Possible Symptoms

 

 

 

 

 

 

 

 

SYMPTOM

 

ACTION

 

 

 

 

 

Output spiking high or raw counts

 

Extend the Blanking (BLK) in increments of 4

 

jumping low

 

 

 

Lower the Threshold

 

 

 

 

 

 

 

 

 

Output spiking low or raw counts

 

Raise the Threshold

 

jumping high

 

Introduce a parallel ground plane

 

 

 

 

 

Output mA erratic

 

Check BLK, THV, GS (section 4.4.2 )

 

 

 

 

 

 

 

 

 

Check for shorts between electronics (+ and – on back) and ground

 

Output greater than 21 mA

 

(case)

 

 

 

 

Change the electronics module

 

 

 

 

Power unit with an isolated supply

 

 

 

 

 

 

Output too low

 

Check for proper supply voltage

 

 

Check terminal strip for leakage to cage.

 

 

 

 

 

 

 

 

 

Raw Counts and Output remain

 

Lower the Threshold

 

constant but in measuring range

 

 

 

Check the probe for obstructions

 

 

 

 

 

 

 

 

 

 

 

 

 

(continued on next page)

 

 

 

 

 

MT2000-0200-1 Rev f (10-2007) DCN0160

19

Image 19
Contents Guided Wave Radar Level Transmitter Table of Contents Appendices List of FiguresIntroduction Measurement cycle consists of the following Ambient TemperatureOverview Storage InformationUltra Low Dielectric ULD Measure Method Direct Reflect Mode standard cont’dMounting Requirements InstallationURV = Probe Length LRV = 0 in. / 0 mm Factory Default ConfigurationShortening of Probe MT2000 Guided Wave Radar GuidelinesMT2000 Guided Wave Radar Guidelines cont’d WiringQuickstart Procedure CommissioningVerify Proper Power-Up Quickstart Procedure cont’dSetting the 4mA and 20 mA Points Using the Pushbutton Menu Calibration Menu LCD Menu OperationThreshold Level Parameter THV and Gain Setting GS Detailed Configuration Parameters Direct Mode Blanking MenuAdvanced Parameter Settings Threshold Level Parameter THV and Gain Setting GS cont’dL1O KO & KG Advanced Parameter Settings cont’d= H * Sqrt DC H + L Ultra Low Dielectric ULD Mode ConfigurationDC = I + H L / H2 DAC TrimFigure Ultra Low Dielectric ULD Mode Configuration cont’dAdvanced Output Configuration Setting the DAC trimBench Calibration of the 4mA & 20mA Points 6 4mA & 20mA Calibration Using Actual Level InputJumper Switch Settings Reversing The Output Action Using Actual Level InputPossible Symptoms Troubleshooting InformationValid Current Loop Outputs Interoperability and Conformance Class Possible Symptoms cont’dElectronics Module Replacement Honeywell DE Output OptionCompatibility Hart Protocol Interface OptionOperating Modes Hart CommunicatorLRV Glossary of TermsAppendix a Sensor Trim AppendicesEND CFG Menu Configuration 2 Hidden MenuTMP To clear the linearization table To access the linearization tableSetting up Linearization Table Points Appendix C LinearizationMT2000 Interface Appendix D Interface ApplicationsChannel B Use and setup of the oscilloscope for the MT2000Scope Settings Channel a PrerequisitesAppendix F CE Certificate of Conformity Appendix G Exida Certificate MT2000-0200-1 Rev f 10-2007 DCN0160 MT2000-0200-01 Rev f 10-2007 DCN0160 MT2000-0200-1 Rev f 10-2007 DCN0160 10b Appendix I Mounting Configurations3a, 3b 10aYear Warranty for Appendix J Warranty Statement