Sanyo MDF-793AT, MDF-593AT Control panel, Break down or power failure at the power supply

Page 13

FREEZER COMPONENTS

Control panel

10

12

11

9

5

 

6

7

8

 

 

1

 

2

3

 

4

 

1.Power switch (POWER): Power all functions except remote alarm and back-up system for AT type.

2.CO2 Back-up switch (CO2 BACK UP): AT type only.

3.CO2 Back-up test switch (TEST) (AT types only): AT type only.

4.Remote alarm switch (REMOTE ALARM): This switch is for remote alarm. In case of operating remote alarm, turn the switch to “ON”.

5.Mode setting key. (PV/SV): This key has functions.

PV: Present freezer compartment temperature is displayed.

SV: Setting temperature is displayed. When the desired temperature is set, this key should be passed (SV side). The temperature setting can be done by using (6),(7),(8), keys, initial setting temperature is -85oC.

6.Digit shift key ( ): The digital of the figure displayed for temperature adjustment can be shifted using this key.

7.Numerical value shift key ( ): The figure that is displayed digitally can be changed by pressing this key.

8.Enter key (ENT): When the desired temperature is reached, press this key. Operation will commence toward this new point.

9.Alarm test key (ALARM): Check that the alarm lamp and the buzzer are functional when the deep-freezer is operating well.

10.Buzzer key (BUZZER): To stop the alarm from sounding, press this key. Should a further abnormality occur, the buzzer will sound automatically.

11.Alarm lamp (ALARM) and buzzer: they operate in the following cases:

a)Break down or power failure at the power supply.

-Alarm lamp flickering, buzzer sounds.

b)Temperature rises about 10oC higher than the setting temperature.

-Alarm lamp flickering, buzzer sounds about 15 minutes later.

12.Filter check lamp (FILTER CHECK): This lamp lights when the condenser filter is clogged. 12

Image 13
Contents MDF-593 MDF-593AT MDF-793 MDF-793AT Contents Symbol shows Precautions for Safe OperationPrevent electric shock or injury Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Control panel cover LockControl panel Break down or power failure at the power supplyInstallation site InstallationAdjust the leveling foot InstallationRemove the packaging materials and tapes Fix the unitSTART-UP of Unit Operation after power failureRemote alarm terminal Temperature settingOperating Instructions Alarm temperature setting Setting of alarm resume time Alarm and safety function Alarm and safety functionsRough scheme Back-up systemFeeding of chart Temperature RecorderReplacement of cell Penholder Setting of back-up temperatureCleaning Routine MaintenanceCleaning of condenser filter Defrosting of inside wallTrouble Shooting Disposal of Unit Location of a nickel-cadmium battery Recycle of NICKEL-CADMIUM BatteryRemoval of nickel-cadmium battery Handling of batteryPerformance SpecificationsSafety check sheet 7FB6P10140500 25 Feb