Sanyo MDF-793AT, MDF-593AT instruction manual Prevent electric shock or injury

Page 4

PRECAUTIONS FOR SAFE OPERATION

WARNING

Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water.

Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire.

Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not adequate, this may result in injury from the unit falling or tipping over.

Never install the unit in a humid place or a place where it is likely to be splashed by water.

Deterioration of the insulation may result which could cause current leakage or electric shock.

Install a leak breaker on the power supply side when you use the unit in humid or wet situations, as well as installing an earth plate. Consult our shops or an electrician for these installations.

Never install the unit in a flammable or volatile location. This may cause explosion or fire.

Never install the unit where acid or corrosive gases are present as current leakage or electric

shock may result due to corrosion.

Use a dedicated power source is used as indicated on the rating label attached to the unit.

Remove dust from the power supply plug before inserting in a power source. A dusty plug or improper insertion may cause a hazard.

Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is not grounded, it will be necessary to install a ground by qualified engineers.

Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such

grounding may cause electric shock in the case of an incomplete circuit.

Never store volatile or flammable substances in this unit. This may cause explosion or fire.

Never store corrosive substances in this unit. This may lead to damage to the inner components or electric parts.

Disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to

prevent electric shock or injury.

3

Image 4
Contents MDF-593 MDF-593AT MDF-793 MDF-793AT Contents Precautions for Safe Operation Symbol showsPrevent electric shock or injury Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Lock Control panel coverBreak down or power failure at the power supply Control panelInstallation Installation siteInstallation Adjust the leveling footRemove the packaging materials and tapes Fix the unitOperation after power failure START-UP of UnitRemote alarm terminal Temperature settingOperating Instructions Alarm temperature setting Setting of alarm resume time Alarm and safety functions Alarm and safety functionBack-up system Rough schemeFeeding of chart Temperature RecorderReplacement of cell Setting of back-up temperature PenholderRoutine Maintenance CleaningCleaning of condenser filter Defrosting of inside wallTrouble Shooting Disposal of Unit Recycle of NICKEL-CADMIUM Battery Location of a nickel-cadmium batteryRemoval of nickel-cadmium battery Handling of batterySpecifications PerformanceSafety check sheet 7FB6P10140500 25 Feb