Sanyo MDF-593AT, MDF-793AT instruction manual Installation site

Page 14

Clean the condenser filter according to page 23 “Cleaning of condenser filter”.

INSTALLATION

Installation site

To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions:

A location not subjected to direct sunlight

The direct sunlight to the freezer causes too high temperature and the chamber temperature may not reach -85oC.

A location with adequate ventilation

Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the refrigeration capacity.

A location away from heat generating sources

Avoid installing the unit near heat-emitting appliances such as gas ranges or stoves. Heat can cause inefficient refrigeration.

A location with a sturdy and level floor

Install the unit on a sturdy floor to avoid vibration and noise. Placing the unit on an unsteady floor may cause vibration and noise.

A location not prone to high humidity

WARNING

Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not adequate, this may result in injury from the unit falling or tipping over.

Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping. Improper installation may result in water spillage or injury from the unit tipping over.

A location not prone to high humidity

WARNING

Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water.

Never install the unit in a humid place or a place where it is likely to be splashed by water.

Deterioration of the insulation may result which could cause current leakage or electric shock.

A location without flammable or corrosive gas

WARNING

Never install the unit in a flammable or volatile location. This may cause explosion or fire.

Never install the unit where acid or corrosive gases are present as current leakage or electric shock

may result due to corrosion.

13

Image 14
Contents MDF-593 MDF-593AT MDF-793 MDF-793AT Contents Precautions for Safe Operation Symbol showsPrevent electric shock or injury Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Lock Control panel coverBreak down or power failure at the power supply Control panelInstallation Installation siteRemove the packaging materials and tapes InstallationAdjust the leveling foot Fix the unitOperation after power failure START-UP of UnitOperating Instructions Temperature settingRemote alarm terminal Alarm temperature setting Setting of alarm resume time Alarm and safety functions Alarm and safety functionBack-up system Rough schemeReplacement of cell Temperature RecorderFeeding of chart Setting of back-up temperature PenholderCleaning of condenser filter Routine MaintenanceCleaning Defrosting of inside wallTrouble Shooting Disposal of Unit Removal of nickel-cadmium battery Recycle of NICKEL-CADMIUM BatteryLocation of a nickel-cadmium battery Handling of batterySpecifications PerformanceSafety check sheet 7FB6P10140500 25 Feb