Sanyo MDF-793AT, MDF-593AT instruction manual Setting of alarm resume time

Page 19

OPERATING INSTRUCTIONS

Setting of alarm resume time

The alarm buzzer is silenced by pressing BUZZER key on the control panel during alarm condition (Initial setting 100: The remote alarm is not silenced).

The buzzer will be activated again after certain suspension if the alarm condition is continued. The suspension time can be set by following the procedure shown in the Table 3 below.

The example in the table is based on the assumption that the desired duration is 30 minutes. Note: The duration is set to 100 (no alarm resume time) at the factory.

Table 3. Setting procedure (change to 30 minutes from no alarm resume time)

 

Description of operation

Key operated

Indication after operation

1

 

-----

The current chamber temperature is

 

 

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

2

Press the numerical value shift key

 

The first digit blinks.

 

 

 

 

 

 

 

 

 

 

for about 5 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set to F25 with the digit shift key

 

The settable digit is shifted.

3

 

 

 

 

 

 

 

When pressed, the figure of settable

 

 

 

 

and numerical value shift key.

 

 

 

 

 

 

 

 

 

 

 

 

 

digit changes.

 

 

 

 

 

 

 

 

 

 

 

4

Press ENT key.

ENT

The current setting is displayed

 

 

 

 

 

 

 

 

and the first digit blinks.

 

 

 

 

 

 

 

 

 

 

 

 

Set to 130 with the digit shift key

 

The settable digit is shifted.

5

 

 

 

 

 

 

 

When pressed, the figure of settable

 

 

 

 

and numerical value shift key.

 

 

 

 

 

 

 

 

 

 

 

 

 

digit changes.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Alarm resume time is memorized

 

 

 

 

6

Press ENT key.

ENT

and the current chamber

 

 

 

 

 

 

 

temperature is displayed.

The settable alarm resume time is between 10 and 60 minutes with 1 minute increment. (The setting is 010, 111, ------- 159, 160). The buzzer would not reset if the resume time is set in 000.

The buzzer is ON when the resume time is changed in F25 while the resume time is counted (the buzzer is silenced by BUZZER key). The count of resume time is stopped.

The setting cannot be changed during power failure.

The set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, any setting before pressing ENT key is not memorized.

It is selectable whether the buzzer is in conjunction with the remote alarm or not when the buzzer is silenced by BUZZER key. (The remote alarm is not in conjunction with the buzzer by factory setting.)

The remote alarm is activated in conjunction with the buzzer if the third digit “1” is set to “0” in the step 5 above

18

Image 19
Contents MDF-593 MDF-593AT MDF-793 MDF-793AT Contents Symbol shows Precautions for Safe OperationPrevent electric shock or injury Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Control panel cover LockControl panel Break down or power failure at the power supplyInstallation site InstallationFix the unit InstallationAdjust the leveling foot Remove the packaging materials and tapesSTART-UP of Unit Operation after power failureRemote alarm terminal Temperature settingOperating Instructions Alarm temperature setting Setting of alarm resume time Alarm and safety function Alarm and safety functionsRough scheme Back-up systemFeeding of chart Temperature RecorderReplacement of cell Penholder Setting of back-up temperatureDefrosting of inside wall Routine MaintenanceCleaning Cleaning of condenser filterTrouble Shooting Disposal of Unit Handling of battery Recycle of NICKEL-CADMIUM BatteryLocation of a nickel-cadmium battery Removal of nickel-cadmium batteryPerformance SpecificationsSafety check sheet 7FB6P10140500 25 Feb