Sanyo MDF-793AT, MDF-593AT instruction manual Alarm and safety functions

Page 20

OPERATING INSTRUCTIONS

Alarm and safety functions

This unit has the alarm and safety functions shown in table below, and also self diagnostic function.

Table 4. Alarm and safety function

Alarm & safety

Situation

Indication

Buzzer

Safety operation

High temperature

When the chamber temp. is

 

Intermittent

 

 

tone with

Remote alarm

approx. 10oC (or 15oC) higher

Alarm lamp blinks.

alarm

than the set temp.

 

12

with 12 min. delay

 

 

min. delay

 

 

 

 

 

Power failure

At power failure

 

Intermittent

 

When power cord is

Alarm lamp blinks.

Remote alarm

alarm

tone

disconnected.

 

 

 

 

 

 

Filter check

When the condenser filter

Filter check lamp lights.

-----

-----

is clogged.

 

 

 

 

 

When there is no key

Chamber temp. is

 

Finishing of each

Auto-return

pressing in each setting

-----

displayed.

setting mode

 

mode for 90 seconds.

 

 

 

 

 

 

If the thermal sensor is

Alarm lamp blinks.

 

 

 

E01 and 50oC is

 

Remote alarm

 

disconnected.

displayed alternately.

Intermittent

 

 

Continuous

 

 

Alarm lamp blinks.

tone

 

If the thermal sensor is

operation

 

E01 and -170oC is

 

 

shirt-circuited.

displayed alternately.

 

 

 

 

 

 

 

If the cascade sensor is

Alarm lamp blinks.

 

 

 

E03 and chamber temp.

 

 

Sensor

disconnected.

 

 

is displayed alternately.

Intermittent

Remote alarm

 

abnormality

If the cascade sensor is

Alarm lamp blinks.

tone

 

 

E04 and chamber temp.

 

 

 

shirt-circuited.

 

 

 

is displayed alternately.

 

 

 

 

 

 

 

If the filter sensor is

Alarm lamp blinks.

 

 

 

E05 and chamber temp.

 

 

 

disconnected.

Intermittent

 

 

is displayed alternately.

Remote alarm

 

 

 

If the filter sensor is

Alarm lamp blinks.

tone

 

 

 

E06 and chamber temp.

 

 

 

short-circuited.

 

 

 

is displayed alternately.

 

 

 

 

 

 

Note

 

 

 

 

The filter check lamp sometimes lights at the start-up under high ambient temperature. The lamp will is off when the chamber temperature is getting lower.

Operation after power failure

The set value (setting of chamber temperature and alarm temperature) is memorized by nonvolatile memory. Accordingly, the freezer resumes the operation with setting before power failure.

19

Image 20
Contents MDF-593 MDF-593AT MDF-793 MDF-793AT Contents Precautions for Safe Operation Symbol showsPrevent electric shock or injury Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Lock Control panel coverBreak down or power failure at the power supply Control panelInstallation Installation siteInstallation Adjust the leveling footRemove the packaging materials and tapes Fix the unitOperation after power failure START-UP of UnitOperating Instructions Temperature settingRemote alarm terminal Alarm temperature setting Setting of alarm resume time Alarm and safety functions Alarm and safety functionBack-up system Rough schemeReplacement of cell Temperature RecorderFeeding of chart Setting of back-up temperature PenholderRoutine Maintenance CleaningCleaning of condenser filter Defrosting of inside wallTrouble Shooting Disposal of Unit Recycle of NICKEL-CADMIUM Battery Location of a nickel-cadmium batteryRemoval of nickel-cadmium battery Handling of batterySpecifications PerformanceSafety check sheet 7FB6P10140500 25 Feb