Sanyo MDF-793AT, MDF-593AT instruction manual

Page 9

CAUTIONS FOR USAGE

This unit is provided with the filter check lamp. If the filter check lamp is flashed, clean the condenser filter. Refer to page 23 “Cleaning of condenser filter”. A clogged filter may cause poor cooling and compressor trouble.

When connecting a liquid CO2 cylinder for the back-up system of AT type freezer, consult your dealer of a neighboring engineering shop of high pressure gas.

Leave the remote alarm switch “OFF” until the inside temperature of the freezer reaches the appropriate level. Leave the back-up system switch “OFF” as well for AT type. After the inside temperature reaches the appropriate level, set the needle for the warning device operation of the temperature recorder.

When the unit is first installed, or re-started after long period of no use, the power failure alarm may not operate, because its Cadnica batteries have completely discharged. This does not denote a malfunction. In this case, continue to operate the unit for about 3 hours. To fully recharge the completely discharged batteries, the unit has to be operated for about 2 days.

Cover the control panel with the control panel cover to avoid the malfunction resulting from an accidental contact with control panel (such as power switch OFF).

8

Image 9
Contents MDF-593 MDF-593AT MDF-793 MDF-793AT Contents Symbol shows Precautions for Safe OperationPrevent electric shock or injury Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Control panel cover LockControl panel Break down or power failure at the power supplyInstallation site InstallationAdjust the leveling foot InstallationRemove the packaging materials and tapes Fix the unitSTART-UP of Unit Operation after power failureTemperature setting Remote alarm terminalOperating Instructions Alarm temperature setting Setting of alarm resume time Alarm and safety function Alarm and safety functionsRough scheme Back-up systemTemperature Recorder Feeding of chartReplacement of cell Penholder Setting of back-up temperatureCleaning Routine MaintenanceCleaning of condenser filter Defrosting of inside wallTrouble Shooting Disposal of Unit Location of a nickel-cadmium battery Recycle of NICKEL-CADMIUM BatteryRemoval of nickel-cadmium battery Handling of batteryPerformance SpecificationsSafety check sheet 7FB6P10140500 25 Feb