Plumbing
Water connection
This machine must be connected to the water supply in accordance with all national and local plumbing codes.
Faucets and drain connections should be situated as close to the machine as possible, and be easily accessible.
All water supply pressures should be between 30 and 145 psi. The machine is supplied with 3 connection hoses
11.5ft. (3.5 m) long with 3/4" (19 mm) female hose thread ends.
If shorter hoses are needed 5 ft. (1.5 m) hoses are supplied with the machine.
Do not shorten hoses!
The machine cannot be operated with an inflow pressure of less than 10 psi flow pressure.
Requirements (refer to the Installation Diagram for more information)
1. Cold water connection:
One 1/2" (13 mm) cold water supply line with a standard 3/4" (19 mm) male hose thread faucet is needed. The cold water supply hose is marked with blue tape.
2. Hot water connection:
One 1/2" (19 mm) hot water supply line with a standard 3/4" male hose thread faucet is needed. The water temperature must not exceed 150 °F / 65 °C (recommended: 140 °F / 60 °C). The hot water supply hose is marked with red tape.
3. DI water connection
One 1/2" (13 mm) deionized water supply line with a standard 3/4" (19 mm) male hose thread faucet is needed. The machine cannot be operated conventionally with a pressure of less than 10 psi flow pressure. In this case the installation of an optional DI pump is required. The DI water supply hose is marked with green tape.