Miele G 7825, G 7826 manual Adding of liquid detergents and neutralizers

Page 25

Adding of liquid detergents and neutralizers

Filling the containers with the relevant detergent or neutralizer

^Turn the lab washer off using the power switch.

^Open the service panel on the bottom right of the machine.

^Take the storage container out of the unit, open it and fill with the agent required. Observe the color code.

^Insert and secure the siphon tube into the container. Observe the color code.

The message clears once the storage containers have been filled.

Please remember to refill the containers in a timely manner, they should not be emptied completely.

If the container is not in use, the refill function for the dosage system not used can be shut off and the respective display message be avoided (see “Program manual - Machine function / Dispensing - Query container").

25

Image 25
Contents Medium capacity Glassware Washer G 7825 / G Page Contents Contents Use Important Safety InstructionsImportant Safety Instructions Important Safety Instructions Disposal of an old appliance Using accessoriesOptions Description of the machineElectronic controls Guide to the machineDisplay Door switch h Start buttonOn/Off button Cursor left Cursor rightUnclean side Guide to the machine Clean side G 7826 only Opening the door Opening and closing the doorTo close the door Mobile unit coding Automatic mobile unit recognition AWKMobile unit coding strip Setting mobile unit codingSet the program place coding Applications Preparing items for cleaningLoading the machine Removing excess soilsApplications Injector mobile unit Laboratory glassware LG Transfer cart for loading and unloading the machine Applications Priming the DOS dispensers Adding of liquid detergents and neutralizersAdditional DOS modules optional Adding of liquid detergents and neutralizers Priming the liquid dispensing systems Dispensing system maintenanceExternal Dispensing systems OptionalOperation Power switchTurning on To change the operating levelFixed program Starting a programFree program selection Turning off Program sequenceAutomatic mobile unit recognition AWK Operating levels B or D To change or cancel a programOperating levels a or C Cancelling a program Serial interface Cleaning the filters in the wash cabinet Cleaning and CareCleaning the fine filter Cleaning the flat filterChanging the coarse filter Drying unit TA Cleaning and careChanging the fine filter Replacing the print paper Maintenance of report printer OptionReplacing the ribbon cartridge Steam/Electro or Electro/Steam conversion InstallationElectrical connection Water connection PlumbingDo not shorten hoses Hot water connectionDrainage Technical Data After sales service Disposal of the packing materials Caring for the environmentDisposal of an old appliance Page All rights reserved