Command :PULS:TRIG[1]:MODE
Long [:SOURce]:PULSe:TRIGger[1]:MODE
Form Set & Query
Parameter CONTinuous | STARt
*RST value STARt
Description Use this command to set/read the TRIGGER OUT generation mode in
pattern mode.
Command :PULS:TRIG[1]:POS
Long [:SOURce]:PULSe:TRIGger[1]:POSition
Form Set & Query
Parameter 1 | 2 | 3 | 4
*RST value 1
Description Use this command to set/read the TRIGGER OUT position in pattern
mode. The specified value selects a segment number for the
Agilent81130A.
Command :PULS:TRIG[1]:VOLT
Long [:SOURce]:PULSe:TRIGger[1]:VOLTage[:LEVel][:IMMediate]
Form Set & Query
Parameter TTL | PECL | SYM | ECLGND | ECLN2V
*RST value TTL