53
Programming Reference
SCPI Instrument Command List
Command :DIG:PATT:SEGM[1|2|3|4]:PRES[1|2]
Long :DIGital[:STIMulus]:PATTern:SEGMent[1|2|3|4]:PRESet[1|2]
Form Set
Parameter <n>,<length>
*RST value Not applicable
Specified Limits <n> 0 to 32768 (integer)
<length> 1 to 65504 (integer)
Description Use this command to set up clock data starting from bit 1 with value 1.
The parameter <n> is used as the divider to generate a CLOCKรทn
sequence (squarewave if NRZ data is selected). The parameter <length>
determines the length of the segment.
n=0 Fill with 0
n=1 Fill with 1
n=2 Sequence = 101010101010101....
n=4 Sequence = 110011001100110....
n=6 Sequence = 111000111000111....
n=8 Sequence = 111100001111000....
and so on.
NOTE The data length meets the same restrictions, than the segment length
(see page 52).
Command :DIG:PATT:SEGM[1|2|3|4]:TYPE[1|2]
Long :DIGital[:STIMulus]:PATTern:SEGMent[1|2|3|4]:TYPE[1|2]
Form Set & Query
Parameter DATA | PRBS | HIGH | LOW
*RST value DATA