Programming Reference

 

 

SCPI Instrument Command List

Description

This command programs the amplitude current of the OUTPUT signal.

 

Note that to set the OUTPUT levels in terms of current, you first have to

 

execute the [:SOURce]:HOLD CURRent command to enable the

 

[:SOURce]:CURRent subsystem.

 

The available current range is limited by the specified voltage limits.

Example

:HOLD CURR

Enable CURRENT subsystem

 

:CURR1 75MA

Set OUTPUT 1 amplitude to 75 mA

Command

:CURR[12]:OFFSet

Long

[:SOURce]:CURRent[12][:LEVel][:IMMediate]:OFFSet

Form

Set & Query

Parameter

Numeric

Parameter suffix

A with engineering prefixes.

*RST value

0.0

A (50 Ω into 50 Ω)

Specified Limits

3.8

V Outputs (50 Ω into short): max. 152 mA typical

 

3.0

V Outputs (50 Ω into short): max. 120 mA typical

Value coupling

Amplitude = High – Low

Offset = High – Low

2

Range coupling

Amplitude

 

Description

This command programs the offset current of the OUTPUT signal. Note

 

that to set the OUTPUT levels in terms of current, you first have to

 

execute the [:SOURce]:HOLD CURRent command to enable the

 

[:SOURce]:CURRent subsystem.

 

The available current range is limited by the specified voltage limits.

Example

:HOLD CURR

Enable CURRENT subsystem

 

:CURR1:OFF 50MA

Set OUTPUT 1 offset to 50 mA

61

Page 61
Image 61
Agilent Technologies 660MHz Execute the SOURceHOLD CURRent command to enable, CURR12OFFSet, ∝A 50 Ω into 50 Ω, Amplitude