Programming Reference

SCPI Instrument Command List

Command

:INIT:CONT

Long

:INITiate:CONTinuous

Form

Set & Query

Parameter

ON OFF 1 0

*RST value

ON

Description

Use this command to enable/disable automatic restart of the instrument

 

(equal to start and stop the instrument). If :ARM:SOURce is set to

 

IMMediate, the value of :INITiate:CONTinuous is ignored.

Command

:CHAN:MATH

Long

:CHANnel:MATH

Form

Set & Query

Parameter

OFF DIGital

*RST value

OFF

Description

Use this command to enable or disable digital channel addition in an

 

instrument with two Output channels installed.

 

With :CHAN:MATH DIGital the digital signals from both channels are

 

”xor’ed” (before the slopes are applied) at OUTPUT 1. The signal of

 

OUTPUT 2 can be used in parallel.

 

This allows you to for example to simulate single or repeated glitches.

Command

:DIG:PATT:LOOP

Long

:DIGital[:STIMulus]:PATTern:LOOP[:LEVel[1]][:COUNt]

Form

Set & Query

Parameter

Numeric

*RST value

1

Specified Limits

1 to 2^20

45