Programming Reference

 

 

 

SCPI Instrument Command List

Command

:CURR[12]:LOW

 

Long

[:SOURce]:CURRent[12][:LEVel][:IMMediate]:LOW

Form

Set & Query

 

 

Parameter

Numeric

 

 

Parameter suffix

A with engineering prefixes.

 

*RST value

–10 mA (50 Ω into 50 Ω)

 

Specified Limits

3.8V Outputs (50 Ω into short): max. 152 mA typical

 

3.0V Outputs (50 Ω into short): max. 120 mA typical

Value coupling

 

 

 

 

Amplitude = High – Low

 

 

Offset =

High – Low

 

 

2

 

 

 

 

Range coupling

High-level

 

 

Description

This command programs the Low-level current of the OUTPUT signal.

 

Note that to set the OUTPUT levels in terms of current, you first have to

 

execute the [:SOURce]:HOLD CURRent command to enable the

 

[:SOURce]:CURRent subsystem.

 

The available current range is limited by the specified voltage limits.

Example

:HOLD CURR

 

Enable CURRENT subsystem

 

:CURR1:LOW 50 MA

Set OUTPUT 1 Low-level to 50 mA

Command

:CURR[12]:LIM

Long

[:SOURce]:CURRent[12]:LIMit[:HIGH]

Form

Set & Query

Parameter

Numeric

Parameter suffix

A with engineering prefixes.

*RST value

+10.0 mA

63

Page 63
Image 63
Agilent Technologies 660MHz manual –10 mA 50 Ω into 50 Ω, 8V Outputs 50 Ω into short max mA typical, High-level, CURR12LIM