Specifications

Agilent 81130A Specifications

Specifications of EXT INPUT/CLK-IN REF Input

Input Parameters

External Input (EXT IN)

External clock/PLL

reference (CLK-IN/REF)

 

 

 

 

 

 

Connectors:

 

SMA(f) 3.5 mm

Termination voltage:

 

–2.10 V to +3.30 V

Termination voltage

 

 

50 mV

resolution:

 

 

 

Input Transitions:

 

 

< 20 ns

Maximum input

 

–3 V to +6 V

voltage:

 

 

 

Threshold:

–1.4 V to +3.7 V

 

ac coupled

Threshold resolution:

50 mV

 

not applicable

Input impedance/

50 Ω typ. / dc

 

50 Ω typ. / ac

coupling:

 

 

 

Input frequency:

0 to 330 MHz

 

External Clock:

 

 

 

170 kHz to 660 MHz

 

 

 

PLL Reference:

 

 

 

1, 2, 5, 10 MHz

Duty cycle:

DC-coupled

 

50% ± 10% duty cycle

Typical delay to

22 ns + 0 ... 1 perioda

 

21 ns

trigger out:

 

 

 

Typical delay to

54 ns + 0 ... 1 perioda

 

53 ns

output:

 

 

 

Sensitivity:

 

< 400 mVpp

 

 

 

 

aThe uncertainty of 1 period can be eliminated if an external clock is used and the following setup and hold times are observed:

setup time: 0.3 ns to 4.3 ns, hold time: –2.8 ns to 4.0 ns.