64
Programming Reference
SCPI Instrument Command List
Description Use this command to set/read the High-level current limit. If you switch
on current limiting, the High-level current cannot be set above t he
programmed limit.
NOTE The current is NOT limited by the OUTPUT hardware, this is a software
limit.
Example :HOLD CURR Enable CURRENT subsystem
:CURR1:LIM 50 MA Set OUTPUT 1 High-level current limit to 50 mA
:CURR1:LIM:STAT ON Switch on OUTPUT 1 limits
Command :CURR[1|2]:LIM:LOW
Long [:SOURce]:CURRent[1|2]:LIMit:LOW
Form Set & Query
Parameter Numeric
Parameter suffix A with engineering prefixes.
*RST value 10.0 mA
Description Use this command to set/read the Low-level current limit. If you switch
on current limiting, the Low-level current cannot be set below the
programmed limit.
NOTE The current is NOT limited by the OUTPUT hardware, this is a software
limit.
Example :HOLD CURR Enable CURRENT subsystem
:CURR1:LIM:LOW -50MA Set OUTPUT 1 Low-level current limit to  50mA
:CURR1:LIM:STAT ON Switch on OUTPUT 1 limits
Command :CURR[1|2]:LIM:STAT
Long [:SOURce]:CURRent[1|2]:LIMit:STATe
Form Set & Query
Parameter ON | OFF | 1 | 0