General Programming Aspects

Status Model

QUEStionable Status Group

Bit

QUEStionable

 

 

0

Voltage warning

1

Current warning

2

Time warning

3

Unused, always 0

4

Unused, always 0

5

Frequency warning

6

Unused, always 0

7

Unused, always 0

8

Unused, always 0

9

Pattern warning

10

Unused, always 0

11

Unused, always 0

12

Unused, always

13

Unused, always 0

14

Unused, always 0

15

Always 0

 

 

The QUEStionable Status group is used to report warning conditions amongst the voltage, current, pulse timing, frequency and pattern parameters. Warnings occur when a parameter, although not outside its maximum limits, could be causing an invalid signal at the output because of the actual settings and uncertainties of related parameters.