24
General Programming Aspects
Status Model
QUEStionable Status GroupThe QUEStionable Status group is used to report warning conditions
amongst the voltage, current, pulse timing, frequency and pattern
parameters. Warnings occur when a parameter, although not outside its
maximum limits, could be causing an invalid signal at the output bec ause
of the actual settings and uncertainties of related parameters.
Bit QUEStionable
0 Voltage warning
1 Current warning
2 Time warning
3 Unused, always 0
4 Unused, always 0
5 Frequency warning
6 Unused, always 0
7 Unused, always 0
8 Unused, always 0
9 Pattern warning
10 Unused, always 0
11 Unused, always 0
12 Unused, always
13 Unused, always 0
14 Unused, always 0
15 Always 0