Programming Reference

SCPI Instrument Command List

Command

:VOLT[12]:LIM:STAT

 

Long

[:SOURce]:VOLTage[12]:LIMit:STATe

Form

Set & Query

 

Parameter

ON OFF 1 0

 

*RST value

OFF

 

Description

This command switches the output limits on or off. When you switch on

 

the output limits cannot program the output-levels beyond the

 

programmed limits, until you switch off the voltage-limits. The limits

 

apply whether you program High/Low levels or Amplitude/Offset levels.

NOTE

You can switch the limits on and off in both the [:SOURce]:CURRent

 

and the [:SOURce]:VOLTage subsystems but the current and voltage

 

limits are not enabled/ disabled independently. The voltage and current

 

limits are always enabled/disabled together.

Example

:HOLD VOLT

Enable VOLTAGE subsystem

 

:VOLT1:LIM 2V

Set OUTPUT 1 High level voltage limit to 2 V

 

:VOLT1:LIM:LOW 0V

Set OUTPUT 1 Low-level voltage limit to 0

 

:VOLT1:LIM:STAT ON

Switch on OUTPUT 1 limits

Command :STATus:OPERation

This command tree accesses the OPERation status group. The OPERation status group is not used by the instrument therefore this command tree is redundant.

:STATus:OPERation[:EVENt]?

:STATus:OPERation:CONDition?

:STATus:OPERation:ENABle

:STATus:OPERation:NTRansition

:STATus:OPERation:PTRansition

81

Page 81
Image 81
Agilent Technologies 660MHz, Agilent 81130A 400 manual Output limits cannot program the output-levels beyond