Programming Reference

SCPI Instrument Command List

Command

:DIG:PATT:LOOP:LENG

Long

:DIGital[:STIMulus]:PATTern:LOOP[:LEVel[1]]:LENGth

Form

Set & Query

Parameter

1 2 3 4

*RST value

1

Description

Use this command to set the number of segments to be repeated within

 

the counted loop.

Example

See previous example (page 47).

Command

:DIG:PATT

Long

:DIGital[:STIMulus]:PATTern[:STATe]

Form

Set & query

Parameter

ON OFF 1 0

*RST value

OFF

Description

Use this command to enable and disable PATTERN mode.

Command

:DIG:PATT:PRBS

Long

:DIGital[:STIMulus]:PATTern:PRBS

Form

Set & Query

Parameter

Numeric

*RST value

7

Specified Limits

7 to 15 (integer)

Description

Use this command to set up PRBS polynom for all PRBS segments on all

 

channels.

48