Specifications

 

Agilent 81130A Specifications

 

Timing Specifications

 

The timing characteristics are measured at 50% amplitude at fastest

 

transitions in continuous mode and 50 Ω load impedance.

NOTE

The Agilent 81130A is designed and recommended for an operation in the

 

frequency range of 170 kHz to 400/660 MHz. However it can be operated

 

in the extended range down to 1 kHz. Changes in specifications below

 

170 kHz are set in brackets [].

 

Period & Frequency

 

Period can also be entered as frequency.

Period & Frequency

Agilent 81130A with

Agilent 81130A with

Agilent 81131A

Agilent 81132A

 

 

 

Period range:

2.5 ns to 1 ms

1.5 ns to 1 ms

Frequency range:

1 kHz to 400.0 MHz

1 kHz to 660.0 MHz

Period/frequency

4 digits, 2 ps best case

resolution:

 

Period accuracya:

± 100 ppm [0.01%]

RMS-jitter:

 

(internal reference,

0.001% + 15 ps

internal clock)

 

 

 

aIn burst mode the first period may be decreased by 150 ps.

Repeatability is typically four times better than accuracy.