Programming Reference

Programming the Instrument Trigger Modes

Burst

Set Burst mode by setting the :TRIGger:COUNt to the burst count required. The trigger source sets the pulse period for the pulses within the burst (See table in “Pulses” on page 39).

:TRIGger:COUNt 16

Burst

of 16 pulse

periods

:TRIGger:SOURce INTernal1

Pulse

period from

internal PLL.

:DIGital:PATTern OFF

Disable pattern data

Pattern

Set Pattern mode by setting the :DIGital[STIMulus]:PATTern:SEGMent[1234]:LENGth to the required pattern length, and switching on digital pattern data. The trigger source sets the pulse period for the data pulses (See table in “Pulses” on page 39):

#Pattern length 512 :DIGital[:STIMulus]:PATTern:SEGMent1:LENGth 512 :DIGital[:STIMulus]:PATTern:SEGMent2:LENGth 0 :DIGital[:STIMulus]:PATTern:SEGMent3:LENGth 0 :DIGital[:STIMulus]:PATTern:SEGMent4:LENGth 0

#Disable counted segment loop :DIGital[:STIMulus]:PATTern:LOOP:COUNt 1

#Jump back to start of segment 1 after the last bit of the last segment (here: segment 1) :DIGital[:STIMulus]:PATTern:LOOP:INFinite[:STATe] ON :DIGital[:STIMulus]:PATTern:LOOP:INFinite:STARt SEGM1

:TRIGger:SOURce INTernal1

Pulse period

from

internal PLL

:DIGital:PATTern ON

Enable

pattern data

:DIGital:SIGNal1:FORMat NRZ

Set OUTPUT 1

data

to NRZ

:ARM:MODE STARted

 

 

 

 

:ARM:SOURce EXT1

Switch

to started

by EXT1

40

Page 40
Image 40
Agilent Technologies Agilent 81130A 400, 660MHz manual Burst, Pattern