Programming Reference

SCPI Instrument Command List

Influence of :TRIGger:COUNt and :TRIGger:COUNt:PULSes[12] in started burst mode:

:ARM:SOUR IMM

TRIG:COUN 1 :ARM:MODE STAR :TRIG:COUN:PULS1 20 :TRIG:COUN 5

:TRIG:COUN:PULS1? 5

:TRIG:COUN? 5

:TRIG:COUN:PULS1 10

:TRIG:COUN? 10

:TRIG:COUN 20

:TRIG:COUN? 10

:TRIG:COUN 8

:TRIG:COUN:PULS1? 8

:TRIG:COUN 1

:TRIG:COUN:PULS1? 8

Set continuous mode

Set Pulse mode

Prepare started mode

Set number of pulses on channel 1 to 20

Set Burst mode with a length of 5 clocks, the number of pulses on both channels will be reduced to 5 if necessary.

Request number of pulses on channel 1

Request number of clock within the started burst

Set number of pulses on channel 1 to 10

Request number of clocks within the started burst

Set number of clocks within the started burst to 20

Request the number of clocks with in the started burst. The return value is 10, because none of the channels will gener- ate more than 10 pulses.

Set the number of clocks within the started burst to 8.

Request the number of pulses on chan- nel 1. The return value is 8, because the number of clocks has been decreased to a value less than the currently used number of pulses on channel 1.

Set Pulse mode

Request the number of pulses on chan- nel 1. The value stays unchanged, since the instrument is no longer in started burst mode.

91

Page 91
Image 91
Agilent Technologies 660MHz, Agilent 81130A 400 manual TRIGCOUNPULS1? ⇒ TRIGCOUN? ⇒