Specifications

Agilent 81130A Specifications

Specification of Trigger Output

This output provides one pulse per period with 50% duty cycle typically. In pattern mode, the trigger pulse can be set to mark the start of any segment.

Trigger Output Specification

Agilent 81130A

 

 

Level (into 50 Ω):

selectable:

 

TTL into GND

 

PECL into +3 V

 

ECL into –2 V

 

ECL into GND

Output impedance:

50 Ω typ.

Trigger pulse width:

50% of period typ.

Maximum external voltage:

–2 V to +3 V

Transition times:

600 ps typ.

Delay from trigger to output:

32 ns typ.

 

 

Output Modes

The output mode determines whether the output signal consists of

•pulses

•bursts of pulses

•patterns of pulses

The output signal is controlled by the Trigger mode.