Programming Reference

SCPI Instrument Command List

Specified limits 0.1 – 99.9%, depends on Width & Period.

Value coupling

Width =

Duty Cycle

Period

100

 

 

 

Description Use this command to program the dutycycle of the pulse signal. If you want to set an absolute pulse-width use [:SOURce]:PULSe:WIDTh[12].

If you want the pulse dutycycle to remain constant when the pulse period is varied (rather than the absolute pulse width use)

[:SOURce]:PULSe:HOLD[12] DCYCle

Example

:PULS:DCYC1

25PCT

Set OUTPUT 1 dutycycle to 25%

 

:PULS:HOLD1

DCYC

Hold dutycycle constant with varying period

Command

:PULS:DEL[12]

Long

[:SOURce]:PULSe:DELay[12]

Form

Set & Query

Parameter

Numeric

Parameter suffix

S with engineering prefixes. You can change the default unit using

 

[:SOURce]:PULSe:DELay[12]:UNIT.

*RST value

0.0

Specified limits

0 to 3.00 µs

Value coupling

 

 

 

 

Phase =

 

Delay

360

 

Period

 

 

 

 

Delay% =

 

Delay

100

 

Period

 

 

 

 

 

68