43
Programming Reference
SCPI Instrument Command List
Command :ARM:LEV[:THR]
Long :ARM[:SEQuence[1] | :STARt][:LAYer]:LEVel[:THReshold]
Form Set & Query
Parameter Numeric
Parameter Suffix V with engineering prefixes.
*RST value +1.0 V
Specified Limits 1.4 V to +3.7 V
Description Use this command to program the triggering threshold of the EXT INPUT
connector.
Example :ARM:LEV 2.5V Set EXT INPUT threshold to 2.5 V
Command :ARM:LEV:TERM
Long :ARM[:SEQuence[1] | :STARt][:LAYer]:LEVel:TERMination
Form Set & Query
Parameter Numeric
Parameter Suffix V with engineering prefixes.
*RST value +0.0 V
Specified Limits 2.1 V to +3.3 V
Description Use this command to program the termination voltage compensation of
the EXT INPUT connector.
Example :ARM:LEV:TERM 1.0V Set EXT INPUT termination voltage to 1.0 V
Command :ARM:MODE
Long :ARM[:SEQuence[1] | :STARt][:LAYer]:MODE
Form Set & Query
Parameter STARted | GATed
*RST value STARted