Programming Reference

 

 

SCPI Instrument Command List

Command

:ARM:LEV[:THR]

 

Long

:ARM[:SEQuence[1] :STARt][:LAYer]:LEVel[:THReshold]

Form

Set & Query

 

Parameter

Numeric

 

Parameter Suffix

V with engineering prefixes.

 

*RST value

+1.0 V

 

Specified Limits

–1.4 V to +3.7 V

 

Description

Use this command to program the triggering threshold of the EXT INPUT

 

connector.

 

Example

:ARM:LEV 2.5V

Set EXT INPUT threshold to 2.5 V

Command

:ARM:LEV:TERM

 

Long

:ARM[:SEQuence[1] :STARt][:LAYer]:LEVel:TERMination

Form

Set & Query

 

Parameter

Numeric

 

Parameter Suffix

V with engineering prefixes.

 

*RST value

+0.0 V

 

Specified Limits

–2.1 V to +3.3 V

 

Description

Use this command to program the termination voltage compensation of

 

the EXT INPUT connector.

 

Example

:ARM:LEV:TERM 1.0V

Set EXT INPUT termination voltage to 1.0 V

Command

:ARM:MODE

Long

:ARM[:SEQuence[1] :STARt][:LAYer]:MODE

Form

Set & Query

Parameter

STARted GATed

*RST value

STARted

43

Page 43
Image 43