Programming Reference

SCPI Instrument Command List

Command

:TRIG:COUN:PULS[12]

 

Long

:TRIGger[:SEQuence[1] :STARt]:COUNt:PULSes[12]

Form

Set & Query

 

Parameter

Numeric

 

*RST value

2

 

Specified limits

2 to 65504

 

Description

Use this command to set/read the number of pulses within a burst at

 

OUTPUT 1 or OUTPUT 2.

 

Examples

To set STARTED BURST of 16 pulse periods and 6 Pulses at

 

Out1, the burst is started by a positive level at the EXT INPUT:

 

:ARM:SOUR EXT1

Set arming from EXT INPUT

 

:ARM:MODE STAR

Set started mode

 

:ARM:SENS POS

Set arming on positive level

 

:TRIG:COUN 16

Burst length 16

 

:TRIG:COUN:PULS1 6

Set 6 pulses at OUTPUT 1

 

:TRIG:SOUR INT1

Pulse period trigger from internal PLL.

 

:DIG:PATT OFF

Disable pattern operating mode

Command

:TRIG:LEV:TERM

 

Long

:TRIGger[:SEQuence[1] :STARt]:LEVel:TERMination

Form

Set & Query

 

Parameter

Numeric

 

Parameter Suffix

V with engineering prefixes.

 

*RST value

0.0 V

 

Specified Limits

–2.1V to +3.3V

 

Description

Use this command to program the termination voltage compensation of

 

the CLK IN connector.

 

Example

:TRIG:LEV:TERM 2.5V

Set CLK IN termination voltage to 2.5 V

92

Page 92
Image 92
Agilent Technologies Agilent 81130A 400, 660MHz manual Output 1 or Output, Triglevterm