Programming Reference

SCPI Instrument Command List

Parameter Suffix

S with engineering prefixes.

 

*RST value

1 s

 

 

 

Specified limits

Agilent 81131A: 2.5 ns to 1 ms

 

 

Agilent 81132A: 1.5 ns to 1 ms

 

Value coupling

 

 

 

 

 

Frequency =

1

 

 

 

 

 

 

 

Period

 

 

 

 

Description

Use this command to set/read the pulse period. Select the pulse period

 

source using :TRIGger:SOURce. The currently selected source is

 

programmed by this command. Note that the specified limits and

 

available resolution depend on the selected source.

 

You cannot set the pulse period if you have selected the CLK IN

 

connector as the frequency source (:TRIG:SOUR EXT2).

Example

:TRIG:SOUR INT

 

 

Select internal PLL as pulse trigger

 

:PULS:PER 25NS

 

 

Set pulse frequency to 25 ns

Command

:PULS:PER:AUTO

 

Long

[:SOURce]:PULSe:PERiod:AUTO

Form

Event

 

Parameter

ONCE

 

*RST value

Not applicable

 

Description

Use this command to measure the period at the CLK IN connector. If the

 

CLK IN connector is the selected pulse period source, you can then read

 

the measured value with

:PULS:PER?

Example

:TRIG:SOUR EXT2

Select ext CLK IN as pulse trigger

 

:PULS:PER:AUTO ONCE

Measure period at CLK IN

 

:PULS:PER?

Query pulse period

71

Page 71
Image 71