Programming Reference

SCPI Instrument Command List

Command

:ROSC:SOUR

 

 

Long

[:SOURce]:ROSCillator:SOURce

Form

Set & Query

 

 

Parameter

INTernal EXTernal

 

*RST value

INT

 

 

Description

Use this command to set/read the reference source for the PLL. If you

 

select the external reference (CLK IN connector) you can choose to use a

 

1 MHz, 2 MHz, 5 MHz or 10 MHz reference signal using

 

:ROSC:EXT:FREQ.

 

 

INTernal

Lock the PLL to its internal reference

 

EXTernal

Lock the PLL to a reference signal at the CLK IN connector. The exter-

 

 

nal reference signal can be 1, 2, 5 or 10 MHz.

Example

:ROSC:SOUR EXT

 

Set external PLL reference (CLK IN)

 

:ROSC:EXT:FREQ 10 MHZ

Set expected PLL reference frequency

 

 

 

to 10 MHz

Command

:ROSC:EXT:FREQ

 

Long

[:SOURce]:ROSCillator:EXTernal:FREQuency

Form

Set & Query

 

Parameter

Numeric

 

*RST value

5 MHz

 

Specified limits

1 MHz, 2 MHz , 5 MHz or 10 MHz

Description

Use this command to set/read the expected reference frequency for the

 

PLL at the CLK IN connector. The external reference can be a 1, 2, 5 or 10

 

MHz signal. Note that if you program any value other than the specified

 

values, the value will be set to the nearest of the specified values.

Example

:ROSC:SOUR EXT

Set external PLL reference (CLK IN)

 

:ROSC:EXT:FREQ 10MHZ

Set expected PLL reference frequency to 10 MHz

76

Page 76
Image 76
Agilent Technologies Agilent 81130A 400, 660MHz manual MHz, 2 MHz, 5 MHz or 10 MHz reference signal using, Roscextfreq