Programming Reference

SCPI Instrument Command List

Specified Limits <n> = 0 to 4 (integer)

Description Use this command to store a complete instrument setting from memory <n> to file filename in the current directory on the memory card.

Memories 1 to 4 are the internal memories. Use memory 0 to store the current instrument setting to a file.

Examples

:MMEM:LOAD:STAT 1,""FREQPERF""

Load FREQPERF into memory 1

 

:MMEM:LOAD:STAT 0,""AMPTEST""

Load AMPTEST as current setting

 

:*SAV

2

Save current setting in memory 2

 

:MMEM:STOR:STAT 2,""SETTING2""

Store memory 2 to file “SETTING2”

 

:*RCL

3

Recall memory 3 as current setting

Command

:OUTP[12]

 

Long

:OUTPut[12][:NORMal][:STATe]

Form

Set & Query

 

Parameter

ON OFF 1 0

 

*RST value

OFF

 

Description

Use this command to switch the normal OUTPUTs on or off.

Example

:OUTP1 ON

Switch on OUTPUT 1

 

:OUTP2 OFF

Switch off OUTPUT 2

Command

:OUTP[12]:COMP

 

Long

:OUTPut[12]:COMPlement[:STATe]

Form

Set & Query

 

Parameter

ON OFF 1 0

 

*RST value

OFF

 

Description

Use this command to switch the complement OUTPUTs on or off.

Example

:OUTP1:COMP ON

Switch on complement OUTPUT 1

 

:OUTP2:COMP OFF

Switch off complement OUTPUT 2

59

Page 59
Image 59
Agilent Technologies 660MHz, Agilent 81130A 400 manual OUTP12COMP, Off