Programming Reference

SCPI Instrument Command List

Command

:CURR[12]:HIGH

 

Long

[:SOURce]:CURRent[12][:LEVel][:IMMediate]:HIGH

Form

Set & Query

 

 

Parameter

Numeric

 

 

Parameter suffix

A with engineering prefixes.

 

*RST value

+10 mA (50 Ω into 50 Ω)

 

Specified Limits

3.8 V Outputs (50 Ω into short): max. 152 mA typical

 

3.0 V Outputs (50 Ω into short): max. 120 mA typical

Value coupling

 

 

 

 

Amplitude = High – Low

 

 

Offset =

High – Low

 

 

2

 

 

 

 

Range coupling

Low-level

 

 

Description

This command programs the High-level current of the OUTPUT signal.

 

Note that to set the OUTPUT levels in terms of current, you first have to

 

execute [:SOURCE]:HOLD CURRent command to enable the

 

[:SOURCE]:CURRent subsystem.

 

The available current range is limited by the specified voltage limits.

Example

:HOLD CURR

 

Enable CURRENT subsystem

 

:CURR1:HIGH 150MA

Set OUTPUT 1 High-level to 150 mA

62

Page 62
Image 62
Agilent Technologies Agilent 81130A 400 +10 mA 50 Ω into 50 Ω, Low-level, Execute Sourcehold CURRent command to enable