62
Programming Reference
SCPI Instrument Command List
Command :CURR[1|2]:HIGH
Long [:SOURce]:CURRent[1|2][:LEVel][:IMMediate]:HIGH
Form Set & Query
Parameter Numeric
Parameter suffix A with engineering prefixes.
*RST value +10 mA (50 into 50 )
Specified Limits 3.8 V Outputs (50 into short): max. 152 mA typical
3.0 V Outputs (50 into short): max. 120 mA typical
Value coupling
Range coupling Low-level
Description This command programs the High-level current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute [:SOURCE]:HOLD CURRent command to enable the
[:SOURCE]:CURRent subsystem.
The available current range is limited by the specified voltage limits.
Example :HOLD CURR Enable CURRENT subsystem
:CURR1:HIGH 150MA Set OUTPUT 1 High-level to 150 mA
Amplitude = High – Low
Offset = High – Low
2