Programming Reference

 

 

SCPI Instrument Command List

Range coupling

Low-level

 

Description

This command programs the High-level voltage of the OUTPUT signal.

 

Note that to set the OUTPUT levels in terms of voltage, you first have to

 

execute the [:SOURce]:HOLD VOLTage command to enable the

 

[:SOURce]:VOLTage subsystem.

 

The available voltage range is limited by the specified current limits.

Example

:HOLD VOLT

Enable VOLTAGE subsystem

 

:VOLT1:HIGH 2V

Set OUTPUT 1 high level voltage to 2 V

Command

:VOLT[12]:LOW

Long

[:SOURce]:VOLTage[12][:LEVel][:IMMediate]:LOW

Form

Set & Query

Parameter

Numeric

Parameter suffix

V with engineering prefixes.

*RST value

–500 mV

Specified Limits

Agilent 81131A: –2.00 V to 3.70 V

 

Agilent 81132A: –2.00 V to 2.90 V

Value coupling

 

 

Amplitude = High – Low

 

 

Offset =

High – Low

 

 

2

 

 

 

 

Range coupling

High-level

 

 

Description

This command programs the Low-level voltage of the OUTPUT signal.

 

Note that to set the OUTPUT levels in terms of voltage, you first have to

 

execute the [:SOURce]:HOLD VOLTage command to enable the

 

[:SOURce]:VOLTage subsystem.

 

The available voltage range is limited by the specified current limits.

Example

:HOLD VOLT

 

Enable VOLTAGE subsystem

 

:VOLT1:LOW 500MV

Set OUTPUT 1 low-level to 500mV

79

Page 79
Image 79
Agilent Technologies 660MHz, Agilent 81130A 400 manual Low-level, VOLT12LOW