79
Programming Reference
SCPI Instrument Command List
Range coupling Low-level
Description This command programs the High-level voltage of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of voltage, you first have to
execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the specified current limits.
Example :HOLD VOLT Enable VOLTAGE subsystem
:VOLT1:HIGH 2V Set OUTPUT 1 high level voltage to 2 V
Command :VOLT[1|2]:LOW
Long [:SOURce]:VOLTage[1|2][:LEVel][:IMMediate]:LOW
Form Set & Query
Parameter Numeric
Parameter suffix V with engineering prefixes.
*RST value 500 mV
Specified Limits Agilent 81131A: 2.00 V to 3.70 V
Agilent 81132A: 2.00 V to 2.90 V
Value coupling
Range coupling High-level
Description This command programs the Low-level voltage of the OUTP UT signal.
Note that to set the OUTPUT levels in terms of voltage, you first have to
execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the specified current limits.
Example :HOLD VOLT Enable VOLTAGE subsystem
:VOLT1:LOW 500MV Set OUTPUT 1 low-level to 500mV
Amplitude = High – Low
Offset = High – Low