Description Use this command to set/read the number of trigger events (pulse periods) to be generated for each arming event in pulse and burst mode (in pattern mode the number of trigger events depends on the used sequence). This corresponds to selecting the event mode on the MODE/ TRG screen:

 

PULSES

Set a trigger count of 1 so that a single pulse period is

 

 

generated for each arming event.

 

BURST of

Set a trigger count of 2 to 65504 so that a burst of 2 to 65504

 

 

pulse periods is generated for each arming event. Switch off

 

 

pattern mode so that a pulse is generated in each pulse

 

 

period. (:DIG:PATT OFF)

NOTE

For a started burst this command will reduce the number of pulses on

 

channel 1 and channel 2 (:TRIGger:COUNt:PULSes[12]) to the value

 

set by :TRIGger:COUNt. Changes of the number of pulses on the

 

channels will increase the value to of :TRIGger:COUNt to reflect the

 

changes on the channels.

 

Examples

To set STARTED BURST of 16 pulse periods and 6 Pulses at

 

Out1, the burst are started by a positive level at the EXT INPUT:

 

:ARM:SOUR EXT1

 

Set arming from EXT INPUT

 

:ARM:MODE STAR

 

Set started mode

 

:ARM:SENS POS

 

Set arming on positive level

 

:TRIG:COUN 16

 

Burst length 16

 

:TRIG:COUN:PULS1 6

Number of pulses at OUTPUT 1

 

:TRIG:SOUR INT1

 

Pulse period trigger from internal PLL.

 

:DIG:PATT OFF

 

Disable pattern operating mode

 

To set GATED PULSES Pulses at Out1, gated by a positive level at the

 

EXT INPUT:

 

 

 

:ARM:SOUR EXT1

 

Set arming from EXT INPUT

 

:ARM:MODE GAT

 

Set arming on levels

 

:ARM:SENS POS

 

Set arming on positive level 1 pulse period

 

:TRIG:COUN 1

 

Single pulse output mode

 

:TRIG:SOUR INT1

 

Pulse period trigger from internal PLL.

 

:DIG:PATT OFF

 

Disable pattern data