Specifications

Pulse Parameter Definitions

Time Reference Point

The time reference point is at the median of the amplitude (50% amplitude point on pulse edge):

100%

50%Median

0%

Pulse Period

The time interval between the leading edge medians of consecutive output pulses:

50%

Pulse Period

Trigger Delay

Interval between trigger point of the external trigger input signal and the trigger output pulse’s leading edge median.

Pulse Width

Interval between leading and trailing edge medians:

50%

Specified Width

The specified and displayed value is that obtained with fastest edges, essentially equal to the interval from the start of the leading edge to the start of the trailing edge. By designing so that the pulse edges turn about their start points, the interval from leading edge start stays unchanged (in