Specifications

Agilent 81130A Specifications

The following rules apply for pattern sequences:

•The resolution of the segment length value depends on the frequency:

Resolution

Frequency in MHz

Period in ns

 

 

 

 

 

16

333.4

... 666.7

1.500 ...

2.999

8

166.7

... 333.3

3.000 ...

5.999

4

83.4 ...

166.6

6.000 ...

11.99

2

41.7 ...

83.3

12.00 ...

23.99

1

min. freq 41.6. ...

24.00 ...

max. period

 

 

 

 

 

•If the counted loop is used, the minimum length of the first segment is twice the resolution.

•An infinite loop over a single segment requires a minimum segment length of twice the resolution.

•Within a segment, PRBS is allowed to be combined with data type PRBS, High or Low only.

•PRBS must be combined with always the same data type in all segments.

Human Interface

Overprogramming

Parameter values can be entered exceeding the specified range.

Warnings and Errors

Warning messages indicate potentially conflicting parameters due to accuracy tolerances.

Error messages indicate conflicting parameters.

108

Page 108
Image 108