General Programming Aspects

Status Model

Status Byte

The status byte summarizes the information from all other status groups. The summary bit for the status byte actually appears in bit 6 (RQS) of the status byte. When RQS is set it generates an SRQ interrupt to the controller indicating that at least one instrument on the bus requires attention. You can read the status byte using a serial poll or *STB?

Bit Description

0

Unused, always 0

1Unused, always 0

2Unused, always 0

3QUESTionable Status Summary Bit

4MAV—Message AVailable in output buffer

5Standard Event Status summary bit

6RQS; ReQuest Service

7OPERation Status summary Bit, unused

Standard Event Status Group

Bit Description

0Operation Complete, set by *OPC

1Unused, always 0

2Query Error

3Device Dependent Error

4Execution Error

5Command Error

6Unused, always 0

7Power On