Programming Reference

SCPI Instrument Command List

Specified Limits Agilent 81131A: –1.95 V to 3.75 V Agilent 81132A: –1.95 V to 2.95 V

Value coupling

High = Offset +

Amplitude

2

 

Low = Offset

Amplitude

 

2

 

 

 

 

Range coupling

Amplitude

 

 

Description

This command programs the offset voltage of the OUTPUT signal. Note

 

that to set the OUTPUT levels in terms of voltage, you first have to

 

execute the [:SOURce]:HOLD VOLTage command to enable the

 

[:SOURce]:VOLtage subsystem.

 

The available voltage range is limited by the specified current limits.

Example

:HOLD VOLT

Enable VOLTAGE subsystem

 

:VOLT1:OFF -800MV

Set OUTPUT 1 offset to –800mV

Command

:VOLT[12]:HIGH

Long

[:SOURce]:VOLTage[12][:LEVel][:IMMediate]:HIGH

Form

Set & Query

Parameter

Numeric

Parameter suffix

V with engineering prefixes.

*RST value

500 mV

Specified Limits

Agilent 81131A: –1.90 V to 3.80 V

 

Agilent 81132A: –1.90 V to 2.50 V

Value coupling

 

Amplitude = High – Low

Offset =

High – Low

2

 

78

Page 78
Image 78
Agilent Technologies Agilent 81130A 400, 660MHz manual Execute the SOURceHOLD VOLTage command to enable, VOLT12HIGH