Agilent Technologies 660MHz manual VOLT12, Agilent 81131A 0.10 Vpp to 3.80 Vpp, Execute

Models: 660MHz Agilent 81130A 400

1 124
Download 124 pages 45.94 Kb
Page 77
Image 77

 

 

 

 

 

Programming Reference

 

 

 

 

 

SCPI Instrument Command List

Command

:VOLT[12]

 

 

 

 

Long

[:SOURce]:VOLTage[12][:LEVel][:IMMediate][:AMPLitude]

Form

Set & Query

 

 

 

 

Parameter

Numeric

 

 

 

 

Parameter suffix

V with engineering prefixes.

*RST value

1.00 V

 

 

 

 

Specified Limits

Agilent 81131A: 0.10 Vpp to 3.80 Vpp

 

Agilent 81132A: 0.10 Vpp to 2.50 Vpp

Value coupling

 

 

Amplitude

 

High = Offset +

 

 

 

 

 

2

 

 

 

 

 

 

 

 

Low = Offset

Amplitude

 

 

 

 

 

2

 

 

 

 

 

 

 

Range coupling

Offset

 

 

 

 

Description

This command programs the amplitude voltage of the OUTPUT signal.

 

Note that to set the OUTPUT levels in terms of voltage, you first have to

 

execute the

[:SOURce]:HOLD VOLTage command to enable the

 

[:SOURce]:VOLTage subsystem.

 

The available voltage range is limited by the specified current limits.

Example

:HOLD VOLT

 

Enable VOLTAGE subsystem

 

:VOLT1 2V

 

Set OUTPUT 1 amplitude to 2 V

Command

:VOLT[12]:OFFSet

Long

[:SOURce]:VOLTage[12][:LEVel][:IMMediate]:OFFSet

Form

Set & Query

Parameter

Numeric

Parameter suffix

V with engineering prefixes.

*RST value

0.0 mV

77

Page 77
Image 77
Agilent Technologies 660MHz VOLT12, Agilent 81131A 0.10 Vpp to 3.80 Vpp, Agilent 81132A 0.10 Vpp to 2.50 Vpp, Execute