Trace Oxygen Analyzer

Installation

 

 

 

4–20 mA DC % Range: Current increases linearly with increasing oxygen, from 4 mA at 0 ppm to 20 mA at full scale ppm. (Full scale = 100% of programmable range)

4–20 mA DC Range ID: 7.2 mA = Low Range

12 mA = Medium Range

16.8 mA = High Range

Table 3-1: Analog Output Connections

Pin Function

3+ Range ID, 4-20 mA, floating

4– Range ID, 4-20 mA, floating

5+ ppm Range, 4-20 mA, floating

6– ppm Range, 4-20 mA, floating

8+ Range ID, 0-1 VDC

23– Range ID, 0-1 V dc, negative ground

24+ ppm Range, 0-1 VDC

7– ppm Range, 0-1 VDC, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the internal alarm relay contacts. Each set of three pins provides one set of Form C relay contacts. Each relay has both normally open and normally closed contact connections. The contact connections are shown in Table 3-2. They are capable of switching up to 3 amperes at 250 VAC into a resistive load. The connectors are:

Threshold Alarm 1:

Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

Can be configured out (defeated).

 

 

 

Teledyne Analytical Instruments

19

Page 31
Image 31