Operation

3000ZA-3X

 

 

 

4.7.2Manual Range Select

-----------------------

Man Rng: R1

-----------------------

Pressing ENTER causes the range R1 or R2 or R3 to start flashing. Then pressing UP or DOWN cycles through the 3 ranges. The selected range is the effective range if the Range Select (above) is set to Man(ual).

4.7.3Range Setpoints

-----------------------

R1: 0-1000.00 ppm

-----------------------

Pressing ENTER causes the value to the right of Rn: to start flashing. Then pressing UP or DOWN adjusts the value. These are the range setpoints for R1, R2, and R3. They may be set to a value in the range of 1.00 ppm to 100% with the stipulation that the each range must have a setpoint equal to or greater than the setpoint of the range below it plus 1ppm.

Teledyne Analytical Instruments

44

Page 56
Image 56
Teledyne 3000ZA-3X operating instructions Manual Range Select, Range Setpoints