Trace Oxygen Analyzer

Maintenance

 

 

 

Maintenance

5.1 Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to replacing the sensors and fuses, and recalibration. For recalibration, see Section 4.4 Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.2 Sensor Replacement

The ZrO2 sensors are sealed transducers with no electrolyte to change or electrodes to clean. When the sensor reaches the end of its useful life, it is replaced. The spent sensor should be discarded according to local regulations.

5.2.1 Removing the Sensor

The ZrO2 sensors is located inside the nylon cell block behind the front panel.

To remove an existing cell (refer to Figure 5-1):

CAUTION: POWER IS APPLIED TO THE INSTRUMENT CIRCUITRY WHENEVER THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE RED SWITCH ON THE FRONT PANEL IS FOR SWITCHING POWER ON OR OFF TO THE DISPLAYS AND OUTPUTS ONLY.

1.Remove power to the instrument by unplugging the power cord at the power source.

 

 

 

Teledyne Analytical Instruments

45

Page 57
Image 57
Teledyne 3000ZA-3X operating instructions Routine Maintenance, Sensor Replacement, Removing the Sensor