Trace Oxygen Analyzer | Maintenance | |
|
|
|
(including the cell) the pressure will be at atmospheric pressure (analyzer vented to atmospheric pressure). Note, all other pressure drops in the sample path are insignificant at these flow rates. This insures that the cell operates at atmospheric pressure. At very high flow rates (off scale of
Example 2: A 3000TA is configured for vacuum service as follows. The unmarked restrictor is placed in the sample vent port. The downstream end of the restrictor is then connected to a vacuum pump and
BY-PASS:
To improve the system response, a
CALIBRATION GAS:
3000 series analyzer requirements for units with
The customer must supply control valves (or restrictors) for any SPAN or ZERO gas source which is attached to the
OPERATION WITHOUT A RESTRICTOR DEVICE:
Operation without a restrictor device is not recommend as mentioned above. A 3000TA without any flow restrictor device was tested on
REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates at various pressures.
|
|
|
Teledyne Analytical Instruments | 59 |