Teledyne 3000ZA-3X Self-Test Routine, Filter Routine, Analog-Out Adjust Routine

Models: 3000ZA-3X

1 72
Download 72 pages 26.68 Kb
Page 46
Image 46

Operation

3000ZA-3X

 

 

 

V1.0

-----------------------

Entering the MODEL screen causes the above screen to be displayed for 3 seconds. There are no modifiable parameters.

4.3.6 Self-Test Routine

-----------------------

Begin Self-Test?

ENTER-YES ESCAPE-NO

-----------------------

Pressing ENTER causes the analyzer to run the same diagnostic that runs on start-up. See section 5.4 for results display.

4.3.7 Filter Routine

-----------------------

Digital Filter ->Index: 5

-----------------------

Pressing ENTER causes the number after 'Index:' to start

flashing. Then pressing UP or DOWN selects a number ranging from 1 to 10. Lower filter values cause the analyzer to converge on new inputs more quickly, but also allow more noise in the sample.

4.3.8 Analog-Out Adjust Routine

-----------------------

4-20mA Offset: 0

Teledyne Analytical Instruments

34

Page 46
Image 46
Teledyne 3000ZA-3X operating instructions Self-Test Routine, Filter Routine, Analog-Out Adjust Routine