Trace Oxygen Analyzer | Maintenance | |
|
|
|
| Atmospheric oxygen | Increase flow rate |
| diffusing through vent | and/or length of vent |
| and altering the | tubing to dilute or |
| oxygen level the | minimize oxygen |
| sensor sees | diffusion from vent to |
|
| sensor. |
Inaccurate zero |
| Turn analyzer off, |
operation (i.e. user has |
| then back on again. |
zeroed the instrument |
| Press the UP and |
using a gas concen- |
| DOWN key |
tration much higher |
| simultaneously while |
than normal for a zero |
| the instrument is |
gas). |
| doing the warm up |
|
| count down. |
|
| Press the ENTER to |
|
| confirm returning the |
|
| analyzer to factor |
|
| defaults. |
|
| |
|
| sensors without doing |
|
| the zero. |
|
| If the analyzer seems |
|
| to be working Ok, up |
|
| to this point, retry to |
|
| zero. |
|
|
|
|
|
|
Teledyne Analytical Instruments | 51 |