Opening and closing the door
The door can only be opened when:
–the lab glassware washer is connected to the electrical supply,
–the red (“emergency
–the
–no program is running.
The door on the clean side (G 7826) can only be opened when:
–DESIN (disinfection) programs have been correctly completed.
or
–the “Door locking" function is active (see “Machine function - Door locking" in the “Programming Manual").
Opening the door
^Press the 5 door button, hold the door handle and open the door.
,Do not touch the heating elements (under the wide area filter) during or directly after the end of a program, you could burn yourself. If the boiler has been programmed to “BOILER READY", be particularly careful of hot water and steam when opening the door. Danger of burning or scalding. The water inlet pipe to the wash cabinet is located, as viewed from the unclean side, underneath the basket runner on the left hand side.
To close the door
^Lift the door upwards and push until it clicks shut.
15