method
1.Unplug the warming unit.2. Unfold the cleaning tool. Wet the foam pads with a nonabrasive solution, such as AlconoxTM brand detergent.
3.Insert the tool from the back of the unit and pull the tool all the way out from the front.
4.Rinse the tool with water and repeat 3 times. Discard the tool according to institutional protocol.
5.Wipe off the unit to remove excess fluid.to clean resistant,
1.Spray a nonabrasive solution inside the slot of the warming unit and let sit for
NOTE: You may use a nonmetal instrument, such as a cotton swab, to clean the upper channels. If you are unable to adequately clean the unit, call 3M Patient Warming technical support.
Storage
Store all components in a cool, dry place when not in use.Call 3M Patient Warming technical support at