Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.www.rangerfluidwarming.com reorder #202524A
English | 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems |
|
|
Model 245
Blood/Fluid Warming System
Preventative Maintenance Checklist
caution
Temperature adjustments cannot be made to the Ranger warming unit. If the test shows the temperature is out of specification, please call 3M Patient Warming technical support at
Tools and equipment
•Thermocouple: 0.005" lead or smaller (part #90020)•Calibrated thermocouple temperature monitor•Aluminum tape•Ranger standard flow blood/fluid warming set•Phillips screwdriver•Saline/waterInlet tube
more or 2mL Drain cassette Insert .remove to cassette from .priming before
0.75"
Aluminum tape
1.80"Thermocouple
Fig. 1: Ranger standard flow blood/fluid warming cassette
12