Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.www.rangerfluidwarming.com reorder #202524A

English

3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

 

 

Model 245

Blood/Fluid Warming System

Preventative Maintenance Checklist

caution

Temperature adjustments cannot be made to the Ranger warming unit. If the test shows the temperature is out of specification, please call 3M Patient Warming technical support at 1-800-733-7775.

Tools and equipment

Thermocouple: 0.005" lead or smaller (part #90020)Calibrated thermocouple temperature monitorAluminum tapeRanger standard flow blood/fluid warming setPhillips screwdriverSaline/water

Inlet tube

more or 2mL Drain cassette Insert .remove to cassette from .priming before

0.75"

Aluminum tape

1.80"Thermocouple

Fig. 1: Ranger standard flow blood/fluid warming cassette