3MTM RangerTM Blood/Fluid and Irrigation Warming Systems

 

 

Troubleshooting

CONDITION

CAUSE

Nothing illuminates on the

• Unit is not turned on, plugged

warming unit panel.

in, or power cord is not plugged

 

into an appropriate outlet.

SOLUTION

Turn unit on. Make sure the power cord is plugged into the power entry module of the warming unit. Make sure the warming unit is plugged into a properly grounded outlet.

CONDITION

Alert indicator illuminates and

alert sounds, alphanumeric display alternately flashes a temperature of 43˚C for the Model 245 (48°C for the Model 247) or higher and the word “HI.”

• Unit failure.

CAUSE

Temporary over-temperature condition because:

An extreme change in flow rates occurred (e.g., from 500 mL/min to stop flow).

Check fuses.

Call 3M Patient Warming technical support.

SOLUTION

Open flow to reduce temperature. Alerts will stop when the display reads 41°C. The unit is ready to use.

Unit was turned on and reached set point temperature before cassette was inserted.

Fluids were prewarmed to above

42°C before being run through the warming unit.

Alerts will stop when the display reads 41°C. The unit is ready to use.

Turn off unit. Discontinue infusion of fluids. Do not warm fluids before infusing them through the Ranger warming unit.

CONDITION

CAUSE

SOLUTION

Alert sounds, alphanumeric

Primary controller failure. Unit

Power to heating plates will shut

display and alert indicator light

will no longer operate.

off if temperature rises to 44°C for

go dark.

 

the Model 245 (50°C for the Model

 

 

247). Turn unit off and unplug it.

 

 

Discontinue use of unit. Discard

 

 

warming set. Alert will continue

 

 

to sound if you do not unplug unit.

 

 

Call 3M Patient Warming technical

 

 

support.