English | 3MTM RangerTM Blood/Fluid and Irrigation Warming Systems |
|
|
Troubleshooting
CONDITION | CAUSE |
Nothing illuminates on the | • Unit is not turned on, plugged |
warming unit panel. | in, or power cord is not plugged |
| into an appropriate outlet. |
SOLUTION
•Turn unit on. Make sure the power cord is plugged into the power entry module of the warming unit. Make sure the warming unit is plugged into a properly grounded outlet.
CONDITION
Alert indicator illuminates and
alert sounds, alphanumeric display alternately flashes a temperature of 43˚C for the Model 245 (48°C for the Model 247) or higher and the word “HI.”
• Unit failure.
CAUSE
Temporary
•An extreme change in flow rates occurred (e.g., from 500 mL/min to stop flow).
•Check fuses.
•Call 3M Patient Warming technical support.
SOLUTION
•Open flow to reduce temperature. Alerts will stop when the display reads 41°C. The unit is ready to use.
•Unit was turned on and reached set point temperature before cassette was inserted.
•Fluids were prewarmed to above
42°C before being run through the warming unit.
•Alerts will stop when the display reads 41°C. The unit is ready to use.
•Turn off unit. Discontinue infusion of fluids. Do not warm fluids before infusing them through the Ranger warming unit.
CONDITION | CAUSE | SOLUTION |
Alert sounds, alphanumeric | Primary controller failure. Unit | Power to heating plates will shut |
display and alert indicator light | will no longer operate. | off if temperature rises to 44°C for |
go dark. |
| the Model 245 (50°C for the Model |
|
| 247). Turn unit off and unplug it. |
|
| Discontinue use of unit. Discard |
|
| warming set. Alert will continue |
|
| to sound if you do not unplug unit. |
|
| Call 3M Patient Warming technical |
|
| support. |
20